S T E R I S R E P L A C E M E N T F I LT E R T E C H N I C A L S P E C I F I C AT I O N S
Whatman Steris ÒAÓ
Whatman Steris ÒBÓ
Replacement Filter
Replacement Filter
Materials of Construction
Entirely of polypropylene
Entirely of polyester
(excluding O-ring)
Filter Media:
Nominal Rated 0.2 µm graded
Absolute Rated1 0.2 µm
density media
Nuclepore track-etch
polyester membrane
Gaskets and O-Rings
Viton
Viton
Biosafety2
Toxicity
USP 23 Class VI Biological Test for Plastics. Conforms to
CFR Title 21.
Endotoxin
<0.25EU/mL as determined using LAL test
Cytotoxicity
Non-Cytotoxic, Non-Reactive
Non-Volatile Residue
<0.5mg per USP Physicochemical Test for Plastics
Water Flow Rate
>6 GPM/psid
>2.15 GPM/psid
(nominal)
(>22 LPM/0.07 bar)
(>8.1 LPM/0.07 bar)
Maximum Operating
75 psid (5.2 bar) at 72¡F (22¡C)
75 psid (5.2 bar) at 72¡F (22¡C)
Pressure
Maximum Operating
176¡F (80¡C) at 5 psi (0.34 bar)
176¡F (80¡C) at 5 psi (0.34 bar)
Temperature
Nominal I.D.
1.0" (2.5 cm)
1.0" (2.5 cm)
Nominal O.D.
2.7" (6.9 cm)
2.7" (6.9 cm)
Surface Area
4.8 ft2 (0.45 m2) per cartridge
11.8 ft2 (1.1 m2) per cartridge
Nominal Lengths
SOE 9.8" (24.9 cm)
SOE 10.8" (27.4 cm)
DOE 9.8" (24.9 cm)
1Rated at >99.99% removal.
2For conditions of the test, please call Whatman, Inc.
Steris ÒAÓ Filter
Steris ÒBÓ Filter
Single Open End
Double Open End
Single Open End
Description
Internal O-Ring
Flat Gasket Seal
2-222 O-Ring
Steris Cat. No.
A1501
Ð
A1504
Whatman Cat. No.
801096
801097
850002
S T E R I S R E P L A C E M E N T F I LT E R S E L E C T I O N I N F O R M AT I O N
Quality System
Certified
Whatman
Whatman is a registered trademark of Whatman International Ltd.
Nuclepore is a registered trademark of Whatman, Inc.
Steris is a registered trademark of Steris Corporation.
12
Please contact
Filter Products at 804-231-4646 to place an order or for more information about this product.